SITRANS FS230 is an ultrasonic clamp-on flowsystem. The SITRANS FST030 transmitter with the proven-in-use FSS200 sensors is ideal for challenging applications and the use in Ex areas. The Sitrans FS230 system is compatible with pipe diameters from DN10 up to DN10.000. For applications in hazardous areas, the external electronics module SITRANS FS DSL is used. It is installed close to the sensors a…
Technical specifications
Enclosure:
Wallmount enclosure: die-cast aluminum, painted, wall or standpipe mounting options, IP65/NEMA 4X for Ex zone 2 or Class 1 Div 2
Industrial enclosure: die-cast aluminum, painted, wall or standpipe mounting options, IP65/NEMA 4X for Ex zone 1/2 or Class 1 Div 1/2
Digital Sensor Link (DSL) – external variant up to 4-path: die-cast aluminum, painted, IP68, for use in Zone 0/1 or Class 1 Div 1/2
Sensors: FSS200 Sensors (Clamp-On) In the universal or high precision version (up to max. 120 ° C / 250 ° F) or high temperature (up to 230°C/446°F), for use in Zone 0/1 or Class 1 Div 1/2
Process values (primary):
Volume flow
Mass flow
Standard volume flow
Flow velocity
Speed of sound
Liquident detection
Totalizer 1, 2 and 3
Communication: HART or Modbus RTU; compatible with SIMATIC PDM; USB service port
Data logger:
4GB SD card, expandable to 32 GB
Data logging in table format (.csv) for: measurement data, alarms, incidents, firmware updates
Logging interval programmable from 1s – 24 hours
I/O options:
3 in/outputs: channel 2, 3 and 4: 4-20 mA active/passive, status output, relay
Inputs channel 5 and 6 on external SITRANS FS DSL: 2 x temperature PT100/500/1000 or 2 x analogue (temperature and pressure)
Display: Full-graphic display, 240 x 160 pixel, backlit
Power supply: 24 V DC / 90 – 240 VAC
Approvals: UL, cUL, CE, ATEX, FM, and more
Measuring accuracy: ≥ 0,3%
Flow range: Liquids ±12 m/s (±40 ft/s), gas, ±40 m/s and more; bidirectional
Pipe diameters: 12,7 mm to 10 m (0.5" to 394")