Application and function
The temperature transmitter AT-3 is applicable to converting resistance of temperature sensor Pt100 to standard current signal 4…20mA. Transmitter has function of compensation non linearity of the sensor. Transmitter can be built in head of sensor. Electrical connection is made with cable cross section up to 1,75 mm2. Transmitter can be ordered with factory set configuration according to customer request. User has also possibility of changing configuration using PC computer with converter USB-AT-3 and special software. In addition to change measuring range user has possibility of configuration behavior of transmitter when sensor is broken and setting compensation of resistance in 2 wire sensor.
Technical data
Sensor type: Pt100 / 2 or 3 wire
Sensor range: -195 ¸ 845°C (18 ¸ 390 W)
Minimum span: 25°C
Output signal: 2 wire 4 ¸ 20 mA current loop
Response time: 1 s
Update time: 500 ms
Power supply (UZ): 10…30 V DC
Load resistance (RO): RO [W] Ł (UZ – 10 V) / 20 mA
Sensor break alarm (configurable): 3,9 mA or 21,5 mA factory setting 21,5 mA
Measurement accuracy: ±0,2°C±0,05% of reading
Thermal drift: ±0,02% /°C (for 20°C)
Loop voltage effect: ±0,2 μA/V
Ambient operating range: -40…+85°C
CÔNG TY TNHH CÔNG NGHIỆP CTMARK VN
Địa chỉ: 119/55A Đường Phan Văn Hùm, Phường Trung Mỹ Tây, Thành phố Hồ Chí Minh, Việt Nam
Mã số thuế: 0317163178
Post code: 70000
Email: info@ctmarkvn.com
www.ctmarkvn.com
Hotline: 0962543178